How to Use Ocular & Stage Micrometer for Calibration
Microscope Calibration Guide
Use an ocular micrometer to measure the specimen image, and use a stage micrometer as the reference scale for calibration. This page explains the basic workflow, calculation method and installation points for practical microscope measurement.
Ocular Micrometer vs. Stage Micrometer — Which to Use?
I want to measure the specimen size under the microscope.
Should I use an Ocular Micrometer ?
or a Stage Micrometer ?
Simply put...
- To measure specimen size under a microscope, use an ocular micrometer.
- A stage micrometer does not measure specimens directly, but is used to check objective lens magnification errors, ensuring more accurate ocular micrometer measurements.
How to Use an Ocular Micrometer
The magnified specimen image is projected onto the ocular micrometer, allowing simultaneous viewing of both the micrometer scale and the sample through the eyepiece. The specimen size is measured by determining the single pitch width on the micrometer scale, which varies depending on the objective lens magnification.
Single pitch width of an ocular micrometer under the microscope =
[Actual width of a single pitch] ÷ [Objective lens magnification]
Example Measurement
The figure below demonstrates measuring the size of a kidney glomerulus.
1. The ocular micrometer : XY11 (10 mm / 100 divisions / 1 pitch = 100 μm),
2. The objective lens : 20x magnification.
The size represented by one pitch on the specimen is 5 μm (= 100 μm ÷ 20).
Since the glomerulus measures 23 pitches horizontally and 22 pitches vertically, the specimen size can be calculated as follows:
Horizontal: 23 × 5 μm = 115 μm
Vertical: 22 × 5 μm = 110 μm

100 μm (1 pitch) ÷ 20 (objective) = 5 μm
X : 40 to 63 : 23 pitch x 5 μm = 115 μm
Y : 40 to 62 : 22 pitch x 5 μm = 110 μm
| Objective Lens Magnification | Magnified Pitch Width 10mm /100 div / pitch 0.1mm |
|---|---|
| 5 x | 20μm (= 100μm ÷ 5) |
| 10 x | 10μm (= 100μm ÷ 10) |
| 20 x | 5μm (= 100μm ÷20) |
| 40 x | 2.5μm (= 100μm ÷ 40) |
| 50 x | 2μm (= 100μm ÷ 50) |
| 100 x | 1μm (= 100μm ÷ 100) |
How to Use an Ocular & Stage Micrometer for Calibration
Objective lenses may have slight magnification errors. A stage micrometer is used to check these errors in advance, so that microscope measurements can be corrected more precisely.
Example of Magnification Error Measurement for Objective Lenses
- Position the micrometers so that the ocular and stage micrometer scales appear parallel when viewed through the eyepiece. This allows for precise magnification error calculation.
- The figure below shows an example using a stage micrometer NOB1 (1 mm / 100 div / pitch = 10 μm) and an ocular micrometer S11 (10 mm / 100 div / pitch = 100 μm) under a 20x objective lens.
[Expected Measurement]
If the magnification is accurate, 10 pitches on the stage micrometer should match 20 pitches on the ocular micrometer:
2000 μm = 10 μm (1 pitch of stage micrometer) × 10 pitches × 20x objective
[Observed Measurement]
However, as seen in this example, if the ocular micrometer reads 21 pitches instead of 20, the actual magnification is slightly higher than 20x, indicating a magnification error.
[Correcting Measurement Values]
To obtain more accurate measurements, multiply the ocular micrometer reading by the error correction factor: 0.95 (= 20 ÷ 21).
Thus, the actual measurement value = ocular micrometer reading × 0.95, ensuring more precise results.
How to Install an Ocular Micrometer
An ocular micrometer is printed in chrome on a glass surface. The printed side is considered the front. Insert it into the eyepiece with the printed side facing the microscope, as shown in the image below.
Each eyepiece model has a different fitting diameter, and compatible micrometer sizes vary depending on the eyepiece manufacturer and model number.
Nikon and Olympus Eyepiece & Ocular Micrometer Compatibility List
How to Identify the Front and Back of an Ocular Micrometer
The scale of the ocular micrometer is printed in chrome on the glass. Once the printed side is identified, place it on the mounting bracket with the printed side facing the objective lens (downward). Refer to the installation instructions above for details. This section explains how to distinguish the front (printed side) and back side of the micrometer.
Frequently Asked Questions
How do I calibrate a microscope using an ocular micrometer and stage micrometer?
Place the stage micrometer on the microscope stage and focus on its scale at the magnification you wish to calibrate. Align the ocular micrometer scale with the stage micrometer divisions and count how many ocular divisions correspond to a known stage micrometer distance. Divide the known distance by the number of ocular divisions to determine the value per ocular division at that magnification.
What is an ocular micrometer and how is it installed?
An ocular micrometer is a glass disc with a scale printed in chrome on one side; it is placed inside the eyepiece of a microscope. Turn the bracket (retaining ring) at the bottom end of the eyepiece counterclockwise and remove it, insert the micrometer into the bracket with the chrome-printed side facing down, then put the bracket back on the eyepiece. The scale appears superimposed on the sample image.
How often should microscope calibration be performed?
Calibration should be performed whenever the objective or eyepiece is changed, when the tube length is adjusted, and at regular intervals (typically annually) for quality assurance. If measurement results seem inconsistent, recalibration is recommended.
Do I need a different calibration for each objective magnification?
Yes. The value of each ocular micrometer division changes with objective magnification. Calibration must be performed separately for each objective (e.g., 4×, 10×, 40×, 100×) and the results recorded. When switching objectives, use the calibration value corresponding to that magnification.